How it works
Take a look
inside Enigma.
Enigma™ brings 1 MHz bipolar radiofrequency (RF), a 36-needle cartridge, and three adjustable parameters into one handheld system. Take a closer look at each component below.

Explore the device
At your fingertips.
The integrated touchscreen displays device status and provides access to dosage, needle depth, and treatment pace. The trigger sits on the pen body.
Adjust all three settings on the pen.
Bipolar radiofrequency
Energy delivered
through the needle array.
Enigma delivers radiofrequency energy through an array of 36 insulated, gold-plated needles, using a 1 MHz bipolar RF system.
The needle array connects to the device through the single-use cartridge. Dosage, needle depth, and treatment pace are selected on the touchscreen by a qualified professional.
View the specifications
Insulated, gold-plated needles
The gliding technique
Enigma QPulse™ Technique
A gliding technique with automatic needle cycling. The needles retract in less than 1/20th of a second. Set dosage, needle depth, and treatment pace on the touchscreen. Refer to current labeling and professional training for operation.
Dosage
10 levels
Select from the device’s ten energy levels. Settings must be determined by a qualified professional.
Depth
0.5–3.5 mm
Needle depth adjusts in 0.25 mm increments across this range.
Pace
Slow. Medium. Fast.
Choose from three pace settings in automatic mode. Changing pace does not change the selected RF energy per pulse. Single-dose mode is also available.
Treatment time
Two passes on the face can be completed in 15–20 minutes.
Timing varies by patient and protocol. Allow additional appointment time for consultation, preparation, and aftercare.
A closer look at dosage
The dose, in numbers.
Each dosage level corresponds to a specified amount of RF energy per pulse, measured in joules (J).
The Operations Manual lists the values shown here. The energy increase between levels is not uniform, so a higher level number does not represent an equal step up in output.
These are device output values. A qualified professional selects settings using current labeling and training.
Request the Operations ManualEnergy per pulse: ±20% tolerance.
Nominal skin impedance: 100–600 Ω.
| Level | Energy per pulse (J) |
|---|---|
| 1 | 0.10 |
| 2 | 0.38 |
| 3 | 0.76 |
| 4 | 1.26 |
| 5 | 1.93 |
| 6 | 2.81 |
| 7 | 3.99 |
| 8 | 5.56 |
| 9 | 7.64 |
| 10 | 10.41 |
Bars share a 0–12 J scale. Values describe energy per pulse, not power in watts or treatment duration.
Source: Enigma RF Operations Manual V21, section 9.1, page 25.
Technical resources
For the details people.
Want to read the Instructions for Use or Operations Manual? Request the current versions from Ellian, along with technical information and training requirements.
Intended use and safety
The Enigma is an FDA-cleared, Class II medical device. 510(k) premarket notification: K262364.
The Enigma RF Device is intended for use in dermatologic procedures for electrocoagulation and hemostasis. Prescription use only. Device use requires appropriate professional licensing, training, and compliance with applicable law.
RF microneedling involves medical risks. The FDA has reported serious complications with certain uses, including burns, scarring, fat loss, disfigurement, and nerve damage. Review the current labeling and FDA safety communication (opens in a new tab). This website does not provide treatment instructions or replace clinical judgment.